前瞻性心电门控技术在双源CT冠状动脉成像中的应用价值
引用本文:陈月芹,孙占国,王林省,史志涛,贾存玮,张谷青,张新东,王彦辉,李慧,王玉红.前瞻性心电门控技术在双源CT冠状动脉成像中的应用价值[J].医学影像学杂志,2012,22(2):178-181.
作者姓名:陈月芹  孙占国  王林省  史志涛  贾存玮  张谷青  张新东  王彦辉  李慧  王玉红
作者单位:济宁医学院附属医院CT室;山东省心血管疾病诊疗重点实验室 山东 济宁 272029
摘    要:目的:比较双源CT冠状动脉成像(CTCA)前瞻性心电门控与回顾性心电门控扫描的辐射剂量及图像质量。方法:以相同纳入标准,选取50例行前瞻性心电门控扫描的患者作为前瞻门控组,另选取50例行回顾性心电门控的患者作为回顾门控组。由两名CT医师分别对两组CTCA图像进行评估。记录并比较两组患者一般临床资料、辐射剂量及图像质量评分。结果:两组患者性别、年龄、体重指数、平均心率、心率波动、扫描范围差异均无统计学意义(P>0.05)。前瞻门控组有效辐射剂量(3.80±0.46)mSv,回顾门控组有效辐射剂量(20.64±3.14)mSv,两组差异具有统计学意义(P<0.001)。两组图像质量的差异无统计学意义(2=1.166,P=0.761)。结论:前瞻性心电门控较回顾性心电门控技术能够在保证图像质量的同时明显降低辐射剂量。

关 键 词:心电门控  冠状动脉成像  体层摄影术  X线计算机  辐射剂量

Application value of prospective electrocardiogram-gating for dual-source CT coronary angiography
CHEN Yue-qin , SUN Zhan-guo , WANG Lin-sheng , SHI Zhi-tao , JIA Cun-wei , ZHANG Gu-qing , ZHANG Xin-dong , WANG Yan-hui , LI Hui , WANG Yu-hong.Application value of prospective electrocardiogram-gating for dual-source CT coronary angiography[J].Journal of Medical Imaging,2012,22(2):178-181.
Authors:CHEN Yue-qin  SUN Zhan-guo  WANG Lin-sheng  SHI Zhi-tao  JIA Cun-wei  ZHANG Gu-qing  ZHANG Xin-dong  WANG Yan-hui  LI Hui  WANG Yu-hong
Affiliation:CT Section,Affiliated Hospital of Jining Medical Collage Shandong Provinicial Key Laboratory of Cardiac Disease Diagnosis and Treament,Jining Shandong 272029,P.R.China
Abstract:Objective:To compare radiation dose and image quality between prospective and retrospective ECG-gating CT coronary angiography(CTCA) with dual-source CT.Methods:A tota l of 100 patients with heart rate less than 70 beats per minute were selected in same standard,50 patients underwent CTCA using prospective ECG-gating as prospective group,another 50 patients underwent CTCA using retrospective ECG-gating as retrospective group.Two independent radiologists assessed the image quality of CTCA used blinding method.General clinical information,scan range,effective radiation dose and image quality were recoded and compered between two group.Results:Sexaul,age,body mass index,heart rate,heart rate variability and scan range between two group had no significant difference(P>0.05).The average effective dose of prospective group and retrospective group was(3.80±0.46) mSv and(20.64±3.14) mSv,respectively(P<0.001).No significant difference of image quality was found between two groups(X2=1.166,P=0.761).Conclusion:Prospective ECG-gating CTCA can effectively reduce radiation dose without compromising image quality.
Keywords:Electrocardiogram gating  Coronary angiography  Tomography  X-ray computed  Radiation dose