螺旋CT颅底区域扫描参数的比较分析
引用本文:刘昌盛,郑小华,谢宝君,童仕平,陈谦学,罗志宏.螺旋CT颅底区域扫描参数的比较分析[J].中国医学影像技术,2003,19(3):360-362.
作者姓名:刘昌盛  郑小华  谢宝君  童仕平  陈谦学  罗志宏
作者单位:1. 武汉大学人民医院放射科CT室,湖北武汉,430060
2. 武汉大学人民医院脑外科,湖北武汉,430060
3. 武汉大学人民医院耳鼻喉科,湖北武汉,430060
摘    要:目的 优化颅底区骨及脑组织螺旋CT扫描及成像参数。方法 对42例疑颅底区病变患者分别行1mm,3mm,5mm层厚螺旋CT扫描,改变扫描及成像参数,以对颞骨内细微结构如听小骨或骨病显示的清晰度及颅底区域脑组织及其病变显示清晰度为标准,分析扫描和成像参数与成像质量的关系。结果 用1mm层厚,小扫描野,0.2mm骨重建,只能非常清晰地显示骨组织及其病变,用3mm层厚,适当增加管电压及照射量,一次扫描,再以1mm间隔,分别以骨和标准算法重建,以骨窗和脑窗显示,能分别清晰显示骨和脑组织及其病变。5mm层厚扫描只能较清晰显示脑组织及病变。结论 颅底区螺旋CT扫描用3mm层厚扫描及成像参数组合的成像质量最佳,可用它来代替常规5mm层厚的扫描技术。

关 键 词:体层摄影术  X线计算机  成像质量  成像参数  颅底  螺旋CT扫描
文章编号:1003-3289(2003)03-0360-03
收稿时间:8/5/2002 12:00:00 AM
修稿时间:2002年8月5日

Analysis on the Parameters of Spiral CT Scanning Basal Brain
LIU Chang-sheng,ZHENG Xiao-hu,XIE Bao-jun,TONG Shi-ping,CHEN Qian-xue and LUO Zhi-hong.Analysis on the Parameters of Spiral CT Scanning Basal Brain[J].Chinese Journal of Medical Imaging Technology,2003,19(3):360-362.
Authors:LIU Chang-sheng  ZHENG Xiao-hu  XIE Bao-jun  TONG Shi-ping  CHEN Qian-xue and LUO Zhi-hong
Affiliation:Department of Radiology, Renmin Hospital of Wuhan University, Wuhan 430060, China;Department of Radiology, Renmin Hospital of Wuhan University, Wuhan 430060, China;Department of Radiology, Renmin Hospital of Wuhan University, Wuhan 430060, China;Department of Radiology, Renmin Hospital of Wuhan University, Wuhan 430060, China
Abstract:Objective To investigate the parameters of scanning or imaging the floor of brain. Methods When head was scanned in collimation of 1mm, 3mm and 5mm. some parameters were changed. As a rule that temporal structures and brain in the floor was viewed clearly, we analyzed relations between some kinds of parameter and imaging quality. Results Bone was clearly viewed in 1mm collimation, small scanning field of volume (SFOV) and 0.2mm bone reconstruction. When the head was scanned in 3mm collimation, high kV and mAs, and standard reconstruction of 1mm interval, bone and brain were clearly viewed perspectively. When it was scanned in collimation of 5mm, only brain was viewed. Conclusion 3mm collimation and other parameters is optimal in the floor scanning.
Keywords:Tomography  X-ray computed  Imaging quality  Brain floor  