首页 | 官方网站   微博 | 高级检索  
     


An intraoperative test device for aortic valve repair
Authors:Ignacio G Berra  Peter E Hammer  Sebastian Berra  Alfredo Oscar Irusta  Seok Chang Ryu  Douglas P Perrin  Nikolay V Vasilyev  Carlos Javier Cornelis  Pablo Garcia Delucis  Pedro J del Nido
Affiliation:1. Cirugia cardiovascular, Hospital Nacional de Pediatria J. P. Garrahan, Ciudad Autonoma de Buenos Aires, Argentina;2. Department of Cardiac Surgery, Boston Children''s Hospital, Boston, Mass;3. Catedra de control, Universidad de Buenos Aires, Ciudad Autonoma de Buenos Aires, Argentina;4. Lew insumos e Innovaciones S.A., Moron, Buenos Aires, Argentina;5. Mechanical Engineering, Texas A&M University, College Station, Tex
Abstract:

Objective

Aortic valve repair is currently in transition from surgical improvisation to a reproducible operation and an option for many patients with aortic regurgitation. Our research efforts at improving reproducibility include development of methods for intraoperatively testing and visualizing the valve in its diastolic state.

Methods

We developed a device that can be intraoperatively secured in the transected aorta allowing the aortic root to be pressurized and the closed valve to be inspected endoscopically. Our device includes a chamber that can be pressurized with crystalloid solution and ports for introduction of an endoscope and measuring gauges. We show use of the device in explanted porcine hearts to visualize the aortic valve and to measure leaflet coaptation height in normal valves and in valves that have undergone valve repair procedures.

Results

The procedure of introducing and securing the device in the aorta, pressurizing the valve, and endoscopically visualizing the closed valve is done in less than 1 minute. The device easily and reversibly attaches to the aortic root and allows direct inspection of the aortic valve under conditions that mimic diastole. It enables the surgeon to intraoperatively study the valve immediately before repair to determine mechanisms of incompetence and immediately after the repair to assess competence. We also show its use in measuring valve leaflet coaptation height in the diastolic state.

Conclusions

This device enables more relevant prerepair valve assessment and also enables a test of postrepair valve competence under physiological pressures.
Keywords:aortic valve repair  device  pressurize  viewer  AR  aortic regurgitation  STJ  sinotubular junction
本文献已被 ScienceDirect 等数据库收录!