首页 | 官方网站   微博 | 高级检索  
     

气体预浓缩-气质联用仪信噪比不确定度评定
引用本文:杨洁.气体预浓缩-气质联用仪信噪比不确定度评定[J].计量与测试技术,2022,49(2):110-112,115.
作者姓名:杨洁
作者单位:自贡检验检测院
摘    要:本文介绍了采用气体预浓缩方法进样的气相色谱-质谱联用仪在环境空气质量监测技术中的应用,以氮气中有机挥发物苯为标准物质,对其信噪比测得值的不确定度进行了评定,当采样量为400mL时,测得峰面积测量重复性为1.1%,信噪比为472,信噪比测得值的相对扩展不确定度为Urel=11%,k=2.

关 键 词:气体预浓缩  气相色谱-质谱联用仪  信噪比  不确定度  

Uncertainty Evaluation of SNR of Chromatography-mass Spectrometry with Pre-concentration of Gas
YANG Jie.Uncertainty Evaluation of SNR of Chromatography-mass Spectrometry with Pre-concentration of Gas[J].Metrology and Measurement Technique,2022,49(2):110-112,115.
Authors:YANG Jie
Abstract:In this paper,the application of Agilent Gas Chromatography-Mass Spectrometry based on pre-concentration of gas in the detection technology of ambient air quality was introduced.And take the organic volatile benzene in nitrogen as a reference material,the uncertainty of signal-to-noise ratio(SNR) measurement result was evaluated,when the sampling volume is 400 mL,the measurement repeatability of peak area was 1.1%,the SNR was 472 and the relative expanded uncertainty of SNR measurement result was Urel=11%,k=2.
Keywords:preconcentration of gas  gas chromatography-mass spectrometry  SNR  uncertainty  benzene
本文献已被 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号