Synthetic single crystal diamond dosimeters for Intensity Modulated Radiation Therapy applications |
| |
Authors: | S Almaviva I Ciancaglioni R Consorti F De Notaristefani C Manfredotti Marco Marinelli E Milani A Petrucci G Prestopino C Verona G Verona-Rinati |
| |
Affiliation: | aINFN-Dipartimento di Ingegneria Meccanica, Università di Roma “Tor Vergata”, Via del Politecnico 1, I-00133 Roma, Italy;bU.O. Fisica Sanitaria, Ospedale San Filippo Neri, Via G. Martinotti 20, 00135 Roma, Italy;cINFN-Dipartimento di Ingegneria Elettronica, Università Roma 3, Via della Vasca Navale 84, 00146 Roma, Italy;dExperimental Physics Department and Center of Excellence NIS, University of Torino 10125, Italy |
| |
Abstract: | A synthetic single crystal diamond Schottky diode, in a p-type/intrinsic/metal structure, deposited by Chemical Vapour Deposition (CVD) and operating in photovoltaic regime, with no external bias voltage applied, was tested as a dosimeter for Intensity Modulated Radiation Therapy (IMRT) applications. The device response was compared with dose measurements from two commercial ionization chambers and a 2D diode array in an IMRT prostate cancer treatment plan. The obtained results indicate that CVD synthetic single crystal diamond-based dosimeters can successfully be used for highly conformed radiotherapy and IMRT dosimetry, due to their small size and high sensitivity per unit volume. |
| |
Keywords: | Single crystal diamond Radiotherapy dosimeter |
本文献已被 ScienceDirect 等数据库收录! |
|