Objective

To assess agreement between a point‐of‐care glucometer (POCG) and a laboratory chemistry analyzer for blood glucose measurements in goats.

Design

Prospective study.

Setting

University teaching hospital.

Animals

Eighteen healthy adult goats.

Investigations

Whole blood samples were obtained via jugular venipuncture prior to premedication with xylazine and butorphanol (T0), following premedication (T20), and after 1 hour of inhalant anesthesia (T60). Each sample was tested with a POCG and a laboratory analyzer (HITA). Agreement was assessed using concordance correlation coefficients and calculation of bias and 95% limits of agreement.

Measurements and Main Results

Mean blood glucose concentration at T0 was 3.9 ± 0.6 mmol/L (70 ± 10 mg/dL; POCG) and 2.9 ± 0.4 mmol/dL (53 ± 8 mg/dL; HITA). Glucose concentrations at T20 were 6.7 ± 2.4 mmol/L (121 ± 43 mg/dL) and 5.4 ± 2.1 mmol/L (97 ± 37 mg/dL) and at T60 were 5.7 ± 1.7 mmol/L (102 ± 31 mg/dL) and 4.7 ± 1.3 mmol/L (85 ± 24 mg/dL) when measured with the POCG and HITA, respectively. The POCG overestimated blood glucose compared to the HITA. The bias ± SD was 1.08 ± 0.53 mmol/L (19.4 ± 9.5 mg/dL) (95% LOA 0.04 to 2.11 mmol/L [0.7 to 38.0 mg/dL]) and the concordance correlation coefficient was 0.82. After correcting the results of the POCG using a mixed‐effects linear model, the bias was 0.0 ± 0.38 mmol/L (0.0 ± 6.8 mg/dL) (95% LOA ± 0.74 mmol/L [± 13.4 mg/dL]) and the concordance correlation coefficient was 0.98.

Conclusions

The POCG overestimated blood glucose concentrations in goats, compared to the HITA, but when the POCG concentrations were corrected, the agreement was excellent.  相似文献   