Application of reflectance spectroscopy to the estimation of uric acid, urea and glucose: an evaluation of the Ames Seralyzer
Authors:JF Stevens and  RG Newall
Affiliation:.Department of Chemical Pathology, Middlesex Hospital Medical School, London; Miles Laboratories, Stoke Poges, Slough
Abstract:An original approach to the measurement of analytes in clinical chemistry has now become available, in which dry reagent strip technology is linked to measurement by reflectance spectroscopy. The present studies have evaluated the performance of the first of these test systems—for uric acid, urea and glucose, in serum—by comparison with conventional liquid chemistry methods. Satisfactory performance in terms of both precision and accuracy was obtained for all three test systems, the current “state-of-the-art” performance criteria being met; the Seralyzer system proved reliable and easy to use in the hands of trained operators. It should find a place as a “Stat” analyser in the laboratory, in specified wards and in Health Centres.